FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2900575 · Received January 7, 2013

Report

Report Number
3004209178-2013-90088
Event Type
Malfunction
Date Received
January 7, 2013
Date of Event
December 20, 2012
Report Date
December 20, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED DISPLACEMENT TEST, REWIND AND BASIC OCCLUSION TEST. THE INSULIN PUMP WAS RECEIVED WITH MOTOR ERROR ALARM STUCK IN BOLUS DELIVERY LOOP. MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED. MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE NOT DETECTED DURING OUR TESTING. UNABLE TO COMPLETE FUNCTIONAL TEST DUE TO MOTOR ERROR ALARM.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE BLOOD GLUCOSE READING WAS 138MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CALLER STATED THAT THE DEVICE WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. ASSISTED THE CUSTOMER TO RUN THE DISPLACEMENT TEST AND THE TEST FAILED. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
6677 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAP

Patients

Seq Age Sex Outcome Treatment
1 26 YR