FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2900517 · Received January 7, 2013

Report

Report Number
2955842-2013-00081
Event Type
Malfunction
Date Received
January 7, 2013
Date of Event
December 11, 2012
Report Date
December 19, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP IS BENT, CAUSING SIDE TO SIDE MISALIGNMENT OF GRIPS. THERE IS A .080 OFFSET AT THE TIPS. THE BENT GRIP DOES NOT EXHIBIT SEPARATION OF THE YAW PULLEY AND PULLEY COVER AT THE GLUE JOINT. ENGINEERING CONCLUDED THAT THE DAMAGE IS LIKELY DUE TO MISHANDLING/MISUSE. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. ENGINEERING EVALUATION ALSO FOUND THAT THE PITCH DOWN CABLE IS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP IS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DOES NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE OTHER CABLES AT WRIST ARE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY PROCEDURE, THE TIPS ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WERE NOT ALIGNED. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
7682 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10121005 881

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURGICAL SYSTEM, ACCESSORIES AND ESU.