ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-00315
- Event Type
- Injury
- Date Received
- January 7, 2013
- Date of Event
- December 10, 2012
- Report Date
- December 11, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.
FOLLOW-UP #1 DATE OF SUBMISSION 04/19/2013 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX REVEALED MULTIPLE "PUMP NOT PRIMED" WARNINGS. A REVIEW OF THE TOTAL DAILY DOSE HISTORY SHOWED THAT THE DAILY INSULIN DELIVERY TOTALS WERE FOUND TO BE WITHIN SPECIFICATION. THE REWIND, PRIME AND LOAD STEPS WERE SUCCESSFULLY PERFORMED WITH NO ALARMS NOTED. THE FORCE SENSOR WAS FOUND TO BE WITHIN CALIBRATION. A 29 HOUR FLOW ACCURACY TEST WAS PERFORMED AND THE PUMP WAS FOUND TO BE DELIVERING WITHIN SPECIFICATIONS. A LOSS OF PRIME WAS INDUCED DURING TESTING AND THE PUMP GAVE THE APPROPRIATE VISUAL AND AUDIBLE ALERT. THE PUMP WAS OPENED AND NO DAMAGE WAS FOUND TO THE FORCE SENSOR CIRCUIT. NO LOSS OF PRIMES OCCURRED DURING TESTING.
ON (B)(6)2012, MOM REPORTS PUMP LOST PRIME AGAIN TODAY AT SCHOOL ABOUT LUNCH TIME AND AS MOM WAS DRIVING TO THE SCHOOL, THE SCHOOL NURSE ATTEMPTED TO REPRIME THE PUMP AND INADVERTENTLY PRIMED 64 UNITS INTO THE PATIENT . MOM DID NOT KNOW THE NURSE HAD PRIMED INSULIN INTO THE PATIENT AND TOOK THE PATIENT HOME: AT HOME, THE BLOOD GLUCOSE (BG) READING BG WAS 58 MG/DL AND THE PATIENT WAS GIVEN JUICE. FIFTEEN MINUTES LATER, AS BG WAS STILL 58 MG/DL , MOM CALLED 911 AND THE PATIENT WAS TRANSPORTED TO THE HOSPITAL, INCOHERENT, WITH BG OF 47 MG/DL AND NOT COHERENT. THE PATIENT WAS ADMITTED AND GIVEN INTRAVENOUS FLUIDS AND IS STILL HOSPITALIZED. PUMP HISTORY SHOWED 9 PRIMES TOTALLING 64 UNITS. THERE WAS ONE LOSS OF PRIME WARNING TODAY. MOM IS INSISTING THE PUMP BE REPLACED. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGATION THAT INADVERTENT INSULIN INFUSION RESULTED IN HOSPITALIZATION FOR HYPOGLYCEMIA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 6343 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 6 YR | Hospitalization |