FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2900230 · Received January 7, 2013

Report

Report Number
1416980-2013-00425
Event Type
Malfunction
Date Received
January 7, 2013
Date of Event
December 1, 2012
Report Date
December 18, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF "INSERT CLAMP ALARM" WAS CONFIRMED DURING DEVICE EVALUATION. THE ROOT CAUSE WAS DUE TO THE SLIDE CLAMP SENSOR BEING OUT OF CALIBRATION. THE SENSOR WAS RECALIBRATED TO CORRECT THE REPORTED CONDITION. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP.

Description of Event or Problem · 1

BAXTER (B)(4) REPORTED A FLOGARD INFUSION PUMP WITH AN INSERT CLAMP ALARM. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT; THEREFORE, NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
6568 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1