FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2900091 · Received January 7, 2013

Report

Report Number
2531779-2013-00284
Event Type
Malfunction
Date Received
January 7, 2013
Report Date
December 9, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION: (B)(6) 2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013, WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX INDICATED REBOOTING HAD OCCURRED. VISUAL INSPECTION REVEALED NO DAMAGE TO THE BATTERY COMPARTMENT OR BATTERY CAP. THE PUMP POWERS ON APPROPRIATELY WITH FUNCTIONAL AUDITORY AND VIBRATORY FEATURES. THE BATTERY CAP SECURES APPROPRIATELY TO THE PUMP WITH NO YELLOW O-RING SHOWING. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER ISSUES OCCURRING. THE COMPLAINT COULD NOT BE DUPLICATED DURING INVESTIGATION.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. THE PUMP REPORTEDLY POWERED OFF UNEXPECTEDLY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
6240 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 17 YR