FDA Adverse Event Injury Summary report: N

PINNACLE MTL INS NEUT36IDX52OD

MDR report key: 2898509 · Received January 7, 2013

Report

Report Number
1818910-2013-10455
Event Type
Injury
Date Received
January 7, 2013
Date of Event
December 11, 2012
Report Date
May 24, 2013
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
KWA
PMA / PMN Number
PK003523
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFO CATALOG AND LOT #. UPDATE: 5/24/2013 - PFS WAS RECEIVED FROM LEGAL, MEDICAL RECORDS WERE RECEIVED FROM LEGAL, AND PART/LOT INFORMATION WAS IDENTIFIED. RECORDS ARE AVAILABLE FOR FURTHER REVIEW. THE INVESTIGATION IS ONGOING. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE AND/OR DHR REVIEW WAS NOT POSSIBLE AS THE PRODUCT AND LOT CODES REQUIRED WERE NOT PROVIDED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE AND/OR DHR REVIEW WAS NOT POSSIBLE AS THE PRODUCT AND LOT CODE REQUIRED WAS NOT PROVIDED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE FOUND NO ADDITIONAL RELATED REPORTS FOR THE LOT CODES 2231746, AY2DJ1000, AND AG1CDA000. A SEARCH OF THE COMPLAINT DATABASE FINDS ADDITIONAL REPORTED INCIDENTS AGAINST LOT CODE 2215523 SINCE ITS RELEASE FOR DISTRIBUTION; HOWEVER, A REVIEW OF THE DEVICE HISTORY RECORDS DID NOT REVEAL ANY RELATED MANUFACTURING ANOMALIES. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

LITIGATION ALLEGES THAT THE PATIENT SUFFERS FROM A HIP SYSTEM THAT DETACHED, DISCONNECTIED, CREATED METALLIC DEBRIS, AND.OR LOOSENED FROM THE PATIENT'S ACETABULUM. PAIN AND DECREASED MOBILITY IS ALSO ALLEGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
6393 PINNACLE MTL INS NEUT36IDX52OD HIP LINER KWA DEPUY ORTHOPAEDICS INC US 2215523

Patients

Seq Age Sex Outcome Treatment
1 Other