FDA Adverse Event Injury Summary report: N

INFUSION PUMP

MDR report key: 2897719 · Received January 4, 2013

Report

Report Number
3007566237-2013-00046
Event Type
Injury
Date Received
January 4, 2013
Date of Event
July 26, 2012
Report Date
December 7, 2012
Manufacturer
MEDTRONIC NEUROMODULATION
Product Code
LKK
PMA / PMN Number
P860004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL EVENT DATE WAS NOT PROVIDED; THIS DATE IS BASED ON THE DATE OF PUBLICATION OF THE ARTICLE. WAS NOT POSSIBLE TO MATCH THIS EVENT TO A PREVIOUSLY REPORTED EVENT. (B)(4).

Description of Event or Problem · 1

KOSTURAKIS, A., GEBHARDT, R. SYNCHROMED II INTRATHECAL PUMP MEMORY ERRORS DUE TO REPEATED MAGNETIC RESONANCE IMAGING. PAIN PHYSICIAN. 2012;15(6):475-477. SUMMARY: CANCER PATIENTS WITH SEVERE REFRACTORY PAIN ARE OFTEN MANAGED WITH IMPLANTABLE DRUG DELIVERY SYSTEMS (IDDS). THE ONLY DRUGS WITH US FOOD AND DRUG ADMINISTRATION APPROVAL FOR INTRATHECAL USE ARE MORPHINE, ZICONOTIDE, AND BACLOFEN. OTHER DRUGS USED AND MIXED INCLUDE, HYDROMORPHONE, BUPIVACAINE, SUFENTANIL, AND FENTANYL. THESE PATIENTS OFTEN UNDERGO MAGNETIC RESONANCE IMAGING (MRI) FOR DISEASE-RELATED MONITORING AND DIAGNOSES. ALTHOUGH UNCOMMON, IDDS CAN FAIL TO RESUME NORMAL FUNCTIONING AFTER MRI, POTENTIALLY CAUSING COMPLICATIONS. THE MAGNETIC FIELD OF AN MRI WILL TEMPORARILY STOP THE ROTOR OF THE PUMP MOTOR AND SUSPEND DRUG DELIVERY FOR THE DURATION OF THE MRI EXPOSURE. THE PUMP SHOULD RESUME NORMAL OPERATION WHEN REMOVED FROM THE MRI MAGNETIC FIELD, BUT THERE IS A POTENTIAL FOR A DELAY IN THE RETURN OF PROPER DRUG INFUSION AND A DELAY IN THE LOGGING OF MOTOR STALL EVENTS AFTER AN MRI IN THE SYNCHROMED II PUMPS. A (B)(6) MAN WHO UNDERWENT MULTIPLE MRIS WITH AN IMPLANTED IDDS EXPERIENCED 2 SEPARATE MEMORY FAILURES LEADING TO MULTIPLE COMPLICATIONS. AFTER THE FIRST PUMP MALFUNCTION, THE PATIENT DEVELOPED WITHDRAWAL SYMPTOMS AND WAS TREATED IN THE EMERGENCY DEPARTMENT. THE FIRST TIME, A MEMORY RESET RESOLVED THE PROBLEM. THE SECOND TIME, 29 MONTHS LATER, THE PATIENT WAS ADMITTED TO THE HOSPITAL TO MANAGE WITHDRAWAL SYMPTOMS AND THE PUMP HAD TO BE EXCHANGED WITH A NEW DEVICE. POST-MRI PUMP INTERROGATION SHOULD BE PERFORMED ON ALL PATIENTS WITH IDDS TO ENSURE PROPER FUNCTIONING OF THE PUMP. SPECIAL ATTENTION SHOULD BE PAID TO PATIENTS RECEIVING BACLOFEN, AS ACUTE WITHDRAWAL CAN BE VERY SERIOUS, EVEN DEADLY. REPORTED EVENT: OVER THE NEXT WEEK FOLLOWING PUMP REPLACEMENT, THE PATIENT EXPERIENCED IRRITATION AND HAD SEROSANGUINEOUS FLUID IN THE SUBCUTANEOUS POCKET SURROUNDING THE INTRATHECAL PUMP THAT FAILED TO RESOLVE WITH ANTIBIOTICS. THE PUMP POCKET WAS REVISED; HOWEVER, DUE TO POOR WOUND HEALING, THE PUMP WAS FINALLY EXPLANTED. ADDITIONAL INFORMATION IS BEING REQUESTED AT THIS TIME; A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
5212 INFUSION PUMP PUMP, INFUSION, IMPLANTED, PROGRAMMABLE LKK MEDTRONIC NEUROMODULATION NEU_UNKNOWN_PUMP

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention CATHETER MODEL: UNK SERIAL/LOT#: UNK