FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 2896794
·
Received December 28, 2012
Report
- Report Number
- 8020893-2012-01572
- Event Type
- Malfunction
- Date Received
- December 28, 2012
- Date of Event
- December 11, 2012
- Report Date
- December 11, 2012
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT AN 840 VENTILATOR STOPPED CYCLING. NO PATIENT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE POWER SUPPLY, ALARM, AND ALARM CABLE. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |