Additional Manufacturer Narrative · 1
(B)(4). THE SIGMA DEVICE EVAL FOUND THE #8 KEY TO BE INOPERABLE. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #8 KEY WILL OCCUR FOLLOWING THE SELECTED KEY'S FUNCTION (EG WHEN THE #1 KEY IS PRESSED 18 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE ABC KEY IS SELECTED, AV WILL BE DISPLAYED INTERFERING WITH THE (B)(4) DRUG SEARCH). SIGMA'S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. AT THIS TIME, THE DATE OF EVENT IS UNK.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE #8 KEY ON A PUMP KEYPAD IS INOPERABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |