FDA Adverse Event Malfunction Summary report: N

NEPTUNE 2 DOCKING STATION (120V)

MDR report key: 2896529 · Received November 20, 2012

Report

Report Number
1811755-2012-04123
Event Type
Malfunction
Date Received
November 20, 2012
Date of Event
September 19, 2012
Report Date
September 19, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A MFR FIELD SERVICE TECH WAS DISPATCHED TO THE USER FACILITY TO EVALUATE THE DEVICE. VISUAL INSPECTION REVEALED BURN MARKS ON THE POWER CORD CONNECTOR. THE SERVICE TECH DETERMINED THAT WATER FROM A FLOOR UTILITY SINK CONTACTED THE POWER MODULE THAT THE POWER CORD CONNECTS WITH, CAUSING AN ELECTRICAL SHORTAGE AND THERMAL DAMAGE TO THE CORD CONNECTOR. THE POWER CORD AND POWER MODULE WERE REPLACED, AND THE DOCKING STATION WAS RETURNED TO USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN EVAL CONDUCTED BY A MFR FIELD SERVICE TECH, THERMAL DAMAGE WAS DISCOVERED ON THE DOCKING STATION POWER CORD CONNECTOR. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE 2 DOCKING STATION (120V) FYD STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK