FDA Adverse Event Malfunction Summary report: N

T4 MULTI STATION BATTERY CHARGER

MDR report key: 2896520 · Received November 20, 2012

Report

Report Number
1811755-2012-04117
Event Type
Malfunction
Date Received
November 20, 2012
Date of Event
October 4, 2012
Report Date
October 30, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYA
PMA / PMN Number
K002348
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVAL, THE CORD WAS FOUND TO HAVE HEAT DAMAGE IN ADDITION TO THE MODULES. THE HEAT DAMAGE ON THE CORD WAS IDENTIFIED AS A POSSIBLE CAUSE OF THE HEAT DAMAGE ON THE MODULES, DUE TO POSSIBLE ARCING WHEN THE CORD WAS PLUGGED INTO THE OUTLET.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS HEAT DAMAGE FOUND ON THE MULTI STATION BATTERY CHARGER WHICH WAS LOCATED IN THE CHARGING ROOM AT THE USER FACILITY. NO PT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 T4 MULTI STATION BATTERY CHARGER FYA STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK