VM6
Report
- Report Number
- 1218950-2012-03827
- Event Type
- Malfunction
- Date Received
- November 20, 2012
- Report Date
- October 23, 2012
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- MHX
- PMA / PMN Number
- K082280
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE CUSTOMER REPORTED THAT WHEN A PT BEING TREATED FOR TACHYCARDIA WAS CONNECTED TO THE VM6 PT MONITOR THE HEART RATE VALUE DISPLAYED AS 140 TO 150 BEATS PER MINUTE. THE STAFF WAS CONSIDERING TREATMENT WITH A BETA-BLOCKER; HOWEVER, THE SPO2 VALUE AND PULSE RATE DID NOT CORRESPOND TO THE DISPLAYED HEART RATE VALUES. THE PT WAS CONNECTED TO A PULSE OXIMETER WHICH SHOWED DIFFERENT VALUES THAN THE MONITOR. NO PT HARM WAS REPORTED. THE MONITOR WAS SHIPPED TO THE PHILIPS SERVICE PROVIDER THIRD PARTY SERVICE CENTER WHERE THE REPAIR TECHNICIAN REPORTED THAT THE MONITOR MAIN BOARD NEEDED TO BE REPLACED. THE REPAIR TECHNICIAN REPLACED THE MONITOR MAIN BOARD TO RESTORE FULL FUNCTION TO THE MONITOR. THIS ISSUE DOES NOT REPRESENT AN EARLY FAILURE, THE MONITOR WAS SHIPPED TO THE CUSTOMER (B)(6) 2007. THE TRENDING DATA DOES NOT SUPPORT THAT THERE IS ANY MANUFACTURING, DESIGN, LABELING, MATERIALS OR SUPPLIER PROBLEM. THE FAILED MAIN BOARD WAS REPLACED TO RESOLVE THE ISSUE AND RESTORE FULL FUNCTION TO THE MONITOR. THE MONITOR WAS REPAIRED AND RETURNED TO THE CUSTOMER. NO FURTHER INVESTIGATION OR ACTION IS WARRANTED.
THE CUSTOMER REPORTED THAT WHEN A PT BEING TREATED FOR TACHYCARDIA WAS CONNECTED TO THE VM6 PT MONITOR THE HEART RATE VALUE DISPLAYED WAS 140 TO 150 BEATS PER MINUTE. NO PT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VM6 | MHX | PHILIPS MEDICAL SYSTEMS | 863064 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |