FDA Adverse Event Injury Summary report: N

CATALYS PRECISION LASER SYSTEM

MDR report key: 2896384 · Received December 28, 2012

Report

Report Number
3005675890-2012-00013
Event Type
Injury
Date Received
December 28, 2012
Date of Event
November 29, 2012
Report Date
December 28, 2012
Manufacturer
OPTIMEDICA CORPORATION
Product Code
OOE
PMA / PMN Number
K121091
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION OF THE INCIDENT INCLUDED THE ANALYSIS OF THE SYSTEM DATABASE, SYSTEM RECORDS, SYSTEM OPTICAL COHERENCE TOMOGRAPHY (OCT) RECORDING, SYSTEM VIDEO DISPLAY RECORDING AND OPERATING ROOM SURGICAL VIDEO. FROM THE ANALYSIS OF THE SYSTEM DATABASE AND SYSTEM RECORDS IT WAS DETERMINED THAT AN INADVERTENT DEVIATION FROM THE DESIGNATED WRITTEN SERVICE PROCEDURE RESULTED IN A MISALIGNMENT OF THE LASER. THE LASER ALIGNMENT ERROR WAS SUBSEQUENTLY CORRECTED PRIOR TO THE NEXT DAY OF SURGERY. THE CAPSULOTOMY WAS REPORTED TO HAVE SOME RESIDUAL ATTACHMENTS. FROM THE ANALYSIS OF THE OPERATING ROOM SURGICAL VIDEO IT WAS DETERMINED THAT THE PHYSICIAN DID NOT USE THE STANDARD CONTINUOUS CURVILINEAR CAPSULORRHEXIS (CCC) SURGICAL TECHNIQUE AND THAT THIS MAY HAVE CAUSED AND/OR CONTRIBUTED TO THE ANTERIOR CAPSULE TEAR. THE SPECIFIC CAUSE(S) OF THE ANTERIOR LENS CAPSULE TEAR ARE UNKNOWN, THE CATALYS SYSTEM OPERATOR MANUAL CONTAINS A WARNING WHICH STATES: "STANDARD CONTINUOUS CURVILINEAR CAPSULORRHEXIS (CCC) SURGICAL TECHNIQUE MUST BE USED FOR SURGICAL REMOVAL OF THE CAPSULOTOMY DISC. THE CAPSULOTOMY MAY HAVE RESIDUAL UNCUT AREAS THAT SHOULD BE COMPLETED BY ADVANCING THE CAPSULE THROUGH THE INCOMPLETELY CUT AREA IN A CIRCUMFERENTIAL FASHION, RATHER THAN PULLING IT RADIALLY. THE USE OF IMPROPER CAPSULOTOMY DISC REMOVAL TECHNIQUE MAY POTENTIALLY CAUSE OR CONTRIBUTE TO ANTERIOR CAPSULE TEAR AND/OR A NONCIRCULAR, IRREGULARLY SHAPED CAPSULOTOMY."

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT WHO UNDERWENT ANTERIOR CAPSULOTOMY, LENS FRAGMENTATION, AND CORNEAL INCISIONS WITH THE CATALYS SYSTEM (SYSTEM) SUBSEQUENTLY EXPERIENCED AN ANTERIOR LENS CAPSULE TEAR SOMETIME DURING THE SURGICAL PROCEDURE TO REMOVE THE CATARACT. THE ANTERIOR CAPSULE TEAR WAS NOTED AFTER REMOVAL OF THE LENS AND CORTEX. IT WAS REPORTED THAT THE PHYSICIAN USED ASPIRATION TO EXTRACT THE CAPSULOTOMY DISC AS OPPOSED TO THE OPTIMEDICA RECOMMENDED CONTINUOUS CURVILINEAR CAPSULORRHEXIS (CCC) CAPSULOTOMY DISC REMOVAL TECHNIQUE. THE SURGEON PLACED THE IOL WITHIN THE LENS BAG AND REMARKED ON THE GOOD POSITION OF THE IOL. NO ADDITIONAL COMPLICATION(S) AND/OR MEDICAL INTERVENTION WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CATALYS PRECISION LASER SYSTEM CATALYS OOE OPTIMEDICA CORPORATION CATALYS-U

Patients

Seq Age Sex Outcome Treatment
1 64 YR Other