ACCU-CHEK SPIRIT COMBO
Report
- Report Number
- 2183996-2012-01931
- Event Type
- Malfunction
- Date Received
- December 5, 2012
- Date of Event
- December 3, 2012
- Report Date
- March 11, 2013
- Manufacturer
- ROCHE INSULIN DELIVERY SYSTEMS INC
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFO CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFO IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
THE COMPLAINT CAN BE VERIFIED. THE MENU BUTTON DOES NOT RESPOND. THERE ARE PARTICLES INSIDE THE HOUSING OF THE MENU BUTTON, AND THESE PARTICLES TEMPORARILY ISOLATE THE AREA OF CONTACT INSIDE THE BUTTON HOUSING. DUE TO THIS PROBLEM, THE MENU BUTTON DOES NOT RESPOND AND IS WITHOUT FUNCTION.
COMPANY REP REPORTED ALL OF THE BUTTONS ON THE INFUSION DEVICE ARE NON-FUNCTIONAL. THIS WAS NOTICED DURING THE START-UP OF THE INFUSION DEVICE. THE INFUSION DEVICE WAS REQUESTED FOR EVAL. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PT DID NOT REQUIRE TREATMENT FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCU-CHEK SPIRIT COMBO | LZG | ROCHE INSULIN DELIVERY SYSTEMS INC | 00700006863 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RELATED ACCESSORIES| INSULIN| INSULIN INFUSION PUMP |