FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2896149
·
Received December 10, 2012
Report
- Report Number
- 1720753-2012-10263
- Event Type
- Malfunction
- Date Received
- December 10, 2012
- Date of Event
- November 29, 2012
- Report Date
- December 10, 2012
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE BOARDS WERE RESEATED AND THE POWER SUPPLY WAS ADJUSTED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM DISPLAYED AN ERROR, LOCKED UP, AND REBOOTED ITSELF. THERE IS NO REPORT OF PT INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |