FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2896134 · Received December 10, 2012

Report

Report Number
1720753-2012-10272
Event Type
Malfunction
Date Received
December 10, 2012
Date of Event
November 20, 2012
Report Date
December 10, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE CONNECTORS WERE RESEATED AND THE FOOT SWITCH WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED AN "X-RAY SWITCH STUCK ERROR" AND WOULD NOT START UP. THERE IS NO REPORT OF PT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1