OT VERIO PRO METER
Report
- Report Number
- 3008382007-2013-00057
- Event Type
- Malfunction
- Date Received
- January 4, 2013
- Report Date
- December 19, 2012
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP (2/15/2013)-DEVICE EVALUATION: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE TEST STRIP HAS PASSED TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2012, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) (B)(4), ALLEGING THE ONETOUCH VERIOPRO METER READ INACCURATELY COMPARED TO ANOTHER DEVICE. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "155 MG/DL" WITH THE SUBJECT METER AND "114 MG/DL" ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS OR SEEK ANY MEDICAL ATTENTION BECAUSE OF THE REPORTED ISSUE. AS THE RESULTS FELL OUTSIDE EXPECTED VALUES FOR METER TO METER ACCURACY TESTING, THIS COMPLAINT IS BEING REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 4396 | OT VERIO PRO METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3299824 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |