OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-00072
- Event Type
- Malfunction
- Date Received
- January 4, 2013
- Report Date
- December 26, 2012
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP (3/19/2013)-DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2012, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) FRANCE ALLEGING THE ONETOUCH VERIO IQ METER READ INACCURATELY COMPARED TO ANOTHER DEVICE. THE PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "HIGH RESULTS AND 206 MG/DL" WITH THE SUBJECT METER AND "143 MG/DL" ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. THE PATIENT DID NOT EXPERIENCE ANY SYMPTOMS OR SEEK ANY MEDICAL ATTENTION BECAUSE OF THE REPORTED ISSUE. AS THE RESULTS FELL OUTSIDE EXPECTED VALUES FOR METER TO METER ACCURACY TESTING, THIS COMPLAINT IS BEING REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 5344 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3387270 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |