FDA Adverse Event Injury Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 2895691 · Received January 3, 2013

Report

Report Number
1031452-2013-00006
Event Type
Injury
Date Received
January 3, 2013
Date of Event
November 26, 2012
Report Date
January 2, 2013
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

IN SPEAKING WITH THE END USER, IT WAS REPORTED SHE THOUGHT SOMETHING WAS ODD ABOUT THE CONCENTRATOR, HAD TRIED TO TURN IT UP DUE TO NEEDING MORE AIR AND SHE WAS UNABLE TO GET IT. SHE TRIED TO TURN IT UP BUT IT WOULD NOT GO OVER 2.5 LITERS AND IT WOULD WORK ITS WAY BACK DOWN. THE END USER HAS HAD THIS CONCENTRATOR FOR ABOUT 3 MONTHS AND NEVER NEEDED MORE AIR BEFORE. SHE DOES NOT KNOW IF THIS HAS BEEN A PROBLEM FROM THE BEGINNING OR NOT. THE END USER HAD HER DAUGHTER CALL AN AMBULANCE TO PICK HER UP DUE TO NOT BEING ABLE TO BREATHE. SHE RECEIVED AIR FROM THE PARAMEDICS. SHE REPORTED HER HEART WAS RUSHING AND HAD PAIN IN HER SIDES DUE TO COUGHING FROM NOT ENOUGH AIR. THE OPERATORS MANUAL INCLUDES THE FOLLOWING: INVACARE RECOMMENDS AN ALTERNATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EVENT OF A POWER OUTAGE, ALARM CONDITION OR MECHANICAL FAILURE. CONSULT YOUR PHYSICIAN OR EQUIPMENT PROVIDER FOR THE TYPE OF RESERVE SYSTEM REQUIRED. THIS EQUIPMENT IS TO BE USED AS AN OXYGEN SUPPLEMENT AND IS NOT CONSIDERED LIFE SUPPORTING OR LIFE SUSTAINING. CONSUMER AFFAIRS HAS TRIED TO LEAVE MESSAGES IN ATTEMPTS TO GATHER INFORMATION; HOWEVER, TO DATE, NO FURTHER INFORMATION HAS BEEN RECEIVED. IT WAS REPORTED THAT THE END USER HAD BEEN PROVIDED CONTACT INFORMATION FOR FURTHER ASSISTANCE WITH THE CONCENTRATOR. PLEASE NOTE THAT ANY FURTHER INFORMATION RECEIVED ON THIS INCIDENT WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

CONSUMER ALLEGES THAT THE PORTABLE OXGEN CONCENTRATOR WOULD NOT GO OVER 2 LITERS WHICH CAUSED HER TO HAVE TO BE HOSPITALIZED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
3349 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC5P

Patients

Seq Age Sex Outcome Treatment
1 62 Hospitalization| O| R