FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2895525 · Received January 3, 2013

Report

Report Number
2955842-2013-00030
Event Type
Malfunction
Date Received
January 3, 2013
Date of Event
November 29, 2012
Report Date
December 6, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE CONDUCTOR WIRE IS BROKEN AT THE YAW PULLEY EXIT. THE WIRE IS DETACHED FROM ITS CONNECTION AT THE GRIP. THE INSTRUMENT FAILED ELECTRICAL CONTINUITY TESTING. THE YAW PULLEY SHOWS NO SIGNS OF ARCING. ENGINEERING EVALUATION ALSO FOUND THAT THE YAW PULLEY IS MISSING A .220 X .060 PIECE OPPOSITE THE CONDUCTOR BREAK, ALLOWING THE GRIP TO MOVE WITHIN THE PULLEY AND HAVE A LARGER RANGE OF MOTION IN YAW. ENGINEERING CONCLUDED THAT IT IS INDETERMINABLE HOW THE PIECE OF YAW PULLEY BROKE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S PROSTATECTOMY PROCEDURE, A WIRE ON THE MARYLAND BIPOLAR FORCEPS INSTRUMENT BROKE. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2589 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 M10110301 102

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SURG SYSTEM, ACCESSORIES AND ESU.