FDA Adverse Event Malfunction Summary report: N

MULTI-LINK RX VISION CORONARY STENT SYSTEM

MDR report key: 2895505 · Received January 3, 2013

Report

Report Number
2024168-2013-00043
Event Type
Malfunction
Date Received
January 3, 2013
Date of Event
December 10, 2012
Report Date
December 10, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
MAF
PMA / PMN Number
P020047
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY:THE DEVICE WAS RETURNED FOR ANALYSIS AND THE REPORTED RESISTANCE WAS NOT CONFIRMED. BASED ON VISUAL, DIMENSIONAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A PROXIMAL RIGHT CORONARY ARTERY. A 3.5 X 12 MM VISION STENT DELIVERY SYSTEM (SDS) WAS BEING ADVANCED THROUGH A 6F GUIDING CATHETER WHEN THERE WAS RESISTANCE NOTED AND THE SDS COULD NOT EXIT THE GUIDING CATHETER. THE SDS WAS REMOVED FROM THE GUIDING CATHETER WITH SOME RESISTANCE. A 3.5 X 8 MM VISION WAS USED SUCCESSFULLY TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2119 MULTI-LINK RX VISION CORONARY STENT SYSTEM CORONARY STENT SYSTEM MAF AV-TEMECULA-CT 2070941

Patients

Seq Age Sex Outcome Treatment
1