FLOGARD
Report
- Report Number
- 1416980-2013-00182
- Event Type
- Malfunction
- Date Received
- January 3, 2013
- Date of Event
- December 13, 2012
- Report Date
- December 13, 2012
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE CAUSE WAS IDENTIFIED TO BE WEAR AND TEAR OF THE PUMP HEAD DOOR. TO CORRECT THIS CONDITION THE PUMP HEAD DOOR WITH REQUIRED PARTS WAS REPLACED AND THE OCCLUSION DETECTORS WERE CALIBRATED AS PER SERVICE MANUAL. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DAMAGED PUMP HEAD DOOR. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS WAS NOT REPORTED BY THE CUSTOMER. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2354 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |