FDA Adverse Event Injury Summary report: N

ASR UNI FEMORAL IMPL SIZE 45

MDR report key: 2895373 · Received January 3, 2013

Report

Report Number
1818910-2013-00311
Event Type
Injury
Date Received
January 3, 2013
Date of Event
November 19, 2012
Report Date
February 25, 2013
Manufacturer
DEPUY INTERNATIONAL
Product Code
KWA
PMA / PMN Number
K040627
Removal / Correction Number
Z-1749/1816-2011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO NEW INFORMATION THAT WOULD CHANGE THE OUTCOME OF THE INVESTIGATION. THE COMPLAINT IS STILL CONSIDERED CLOSED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE REPORTED DEVICES WAS NOT POSSIBLE AS THEY WERE NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES FINDS NO OTHER REPORTS AGAINST THE STEM/SLEEVE PRODUCT AND LOT CODE COMBINATIONS SINCE THEIR RELEASE TO DISTRIBUTION. THE INVESTIGATION CAN DRAW NO CONCLUSIONS WITH THE INFORMATION PROVIDED. IT SHOULD BE NOTED THAT THE REPORTED CORROSION WAS SAID TO HAVE BEEN WIPED OFF IN SURGERY. THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Description of Event or Problem · 1

PATIENTS REVISION OPERATIVE RECORDS FOR THE LEFT HIP WERE RECEIVED. RECORDS INDICATE UPON REVISION ATYPICAL LOCAL TISSUE REACTION, A SMALL PSEUDOTUMOR, A LARGE GREENISH BROWN BURSA, AND A LARGE DOME CYST.

Description of Event or Problem · 1

LITIGATION PAPERS ALLEGE THAT THE BILATERAL PATIENT SUFFERED FROM HIP PAIN, BACK AND GROIN PAIN, INABILITY TO STAND FOR LONG PERIODS OF TIME, DIFFICULTY WALKING, WALKED WITH A LIMP, BLURRINESS AND VISION PROBLEMS AND WAS INJURED BY EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HER BLOOD. SHE HAD A COLLECTION OF FLUIDS IN HER LEFT HIP JOINT WHICH, ALONG WITH HER CONSTANT PAIN AND ELEVATED ION LEVELS, LED TO REVISION SURGERY OF THE LEFT HIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2129 ASR UNI FEMORAL IMPL SIZE 45 TOTAL HIP REPLACEMENT KWA DEPUY INTERNATIONAL 2454974

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention