FDA Adverse Event Malfunction Summary report: N

PERMANENT CAUTERY SPATULA INSTRUMENT

MDR report key: 2895353 · Received January 3, 2013

Report

Report Number
2955842-2013-00025
Event Type
Malfunction
Date Received
January 3, 2013
Date of Event
November 30, 2012
Report Date
December 5, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS RETURNED WITH A SECTION OF THE CERAMIC SLEEVE BROKEN OFF. THE SLEEVE IS MISSING PIECES AT THE DISTAL END WITH AN OVERALL FOOTPRINT MEASURING ROUGHLY .090 X .090. SLEEVE EXHIBITS VARIOUS SCRATCH MARKS BELOW THE BROKEN SECTION. SPATULA ALSO EXHIBITS LIGHT SCRATCH MARKS IN THE AXIAL DIRECTION, INDICATING POSSIBLE ATTEMPT AT SCRAPING OFF BIODEBRIS, WHICH MAY HAVE DAMAGED THE CERAMIC. ENGINEERING CONCLUDED THAT THE DAMAGE WAS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. O DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE TIP ON THE PERMANENT CAUTERY SPATULA INSTRUMENT CRUMBLE AND BROKE LOOSE, WHILE THE SURGEON WAS PERFORMING LYMPH NODE DISSECTION. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
3592 PERMANENT CAUTERY SPATULA INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420184-06 M10120404 312

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES