FDA Adverse Event Summary report: N

OT VERIO METER

MDR report key: 2895111 · Received January 3, 2013

Report

Report Number
3008382007-2013-00021
Date Received
January 3, 2013
Report Date
December 18, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K093745
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1: ((B)(6) 2013)-DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED (B)(4) 2013 AND EVALUATED BY PRODUCT ANALYSIS (PA) ON (B)(4) 2013, WITH THE FOLLOWING FINDINGS: THE METER WAS RETURNED WITHOUT THE BATTERIES. AFTER PA INSERTED BATTERIES, THE METER WAS EVALUATED. THE METER PASSED TESTING WITH NO FAULTS FOUND. THE METER WAS FOUND TO FUNCTION PROPERLY. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

PRODUCTS REPLACED AND REQUESTED BACK FOR INVESTIGATION.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING INACCURATE RESULTS. THE PATIENT OBTAINED THE FOLLOWING RESULTS BACK TO BACK 16. 6, 11.5, 9.4 MMOL/L. THIS COMPLAINT IS BEING REPORTED BECAUSE, THE REPORTED RESULTS DID NOT MEET LIFESCAN'S ACCURACY/PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2613 OT VERIO METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3355888

Patients

Seq Age Sex Outcome Treatment
1 60 YR