V.A.C. THERAPY
Report
- Report Number
- 3009897021-2012-00005
- Event Type
- Injury
- Date Received
- December 27, 2012
- Date of Event
- May 30, 2012
- Report Date
- October 26, 2012
- Manufacturer
- KCI USA INC.
- Product Code
- OMP
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- NURSE
Narratives
ON (B)(4) 2012, KCI FIELD SERVICE TESTED THE DEVICE PER QUALITY CONTROL (QC) PROCEDURE AND DETERMINED THE UNIT MET SPECIFICATIONS PRIOR TO PATIENT PLACEMENT, WHICH INCLUDED THE PUMP AND ALARM FUNCTIONS. AS A RESULT OF THE ALLEGATION THAT THE UNIT WAS NOT FUNCTIONING PROPERLY, THE UNIT WAS REMOVED FROM THE HOSPITAL BY KCI AND REPLACED WITH ANOTHER UNIT. THE DEVICE WAS RETURNED TO KCI FOR EVALUATION. EVALUATION OF THE DEVICE IDENTIFIED A PROBLEM WITH A TRANSDUCER CONNECTION WHICH CAN POTENTIALLY CAUSE IMPAIRED PUMP FUNCTIONALITY, LIKELY CAUSED BY THE UNIT BEING DROPPED DURING USE. THE KCI EVALUATION ALSO DETERMINED THE IDENTIFIED TRANSDUCER PROBLEM WOULD NOT AFFECT THE OPERATION OF THE ALARM SYSTEM AND THAT IF THE PUMP WAS IMPAIRED, THE APPROPRIATE ALARM WOULD ALERT THE USER IMMEDIATELY. ON (B)(4) 2012, KCI FIELD SERVICE TESTED THE DEVICE PER QUALITY CONTROL (QC) PROCEDURE AND DETERMINED THE UNIT MET SPECIFICATIONS POST REPAIR.
ON (B)(6) 2012, THE FOLLOWING INFORMATION WAS OBTAINED FROM THE NURSE: THE NURSE ALLEGED THE ACTIV.A.C. UNIT WAS NOT FUNCTIONING PROPERLY AS THE NURSE REPORTEDLY OBSERVED NO SUCTION AT THE DRESSING. ON (B)(6) 212, THE FOLLOWING INFORMATION WAS OBTAINED FROM THE NURSE: THE NURSE STATED THE ACTIV.A.C. UNIT DISPLAYED 150 MMHG (TARGET PRESSURE) BUT ALLEGED THE GRANUFOAM DRESSING WAS SOFT, WHICH SHE BELIEVED TO BE INDICATION NEGATIVE PRESSURE WAS NOT PRESENT AT THE DRESSING. UPON REMOVING THE PATIENT'S DRESSING, THE NURSE NOTED LESS THAN EXPECTED EXUDATES. THE NURSE SUCTIONED APPROXIMATELY 200 CC OF GREEN DRAINAGE FROM THE WOUND BED AND OBSERVED MACERATION AND A SLOUGH LIKE FILM COVERING THE WOUND BED. LABORATORY TESTING WAS PERFORMED AND THE PATIENT HAD ELEVATED C-REACTIVE PROTEIN AND WHIT BLOOD CELL COUNT LEVELS. THE PATIENT WAS PLACED ON ANTIBIOTICS AND THE PATIENT'S CONDITIONED RESOLVED. IT WAS REPORTED THAT THE WOUND HEALED COMPLETELY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | V.A.C. THERAPY | OMP | KCI USA INC. | ACTIV.A.C. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Required Intervention |