FDA Adverse Event Malfunction Summary report: N

UBIT-IR300 INFRARED SPECTROPHOTOMETER

MDR report key: 2894963 · Received November 13, 2007

Report

Report Number
3005622096-2007-00001
Event Type
Malfunction
Date Received
November 13, 2007
Date of Event
October 2, 2007
Report Date
November 9, 2007
Manufacturer
OTSUKA ELECTRONICS CO. LTD.
Product Code
JJQ
PMA / PMN Number
K013371
Removal / Correction Number
07-058-CMD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SUMMARY OF PRELIMINARY INVESTIGATION OF DEVICE INVOLVED - INVESTIGATION ONGOING: THE DEVICE INVOLVED, UBIT-IR300 SPECTROPHOTOMETER, SERIAL NO (B)(4), WAS SHIPPED TO EXPONENT FAILURE ANALYSIS ASSOCIATES ((B)(4)) FOR ANALYSIS AND CAUSE DETERMINATION. OTSUKA ELECTRONICS CO LTD PERSONNEL TRAVELED FROM (B)(4) TO (B)(4) TO MEET WITH EXPONENT AND REVIEW RESULTS TO DATE. A VISUAL INSPECTION OF THE INSTRUMENT SHOWED NO SIGNS OF DAMAGE ON THE EXTERIOR OR INTERIOR OF THE UNIT WHEN THE OUTER COVER WAS REMOVED. ALL VISUAL DAMAGE WAS CONFINED TO THE INSIDE OF THE POWER SUPPLY. THE POWER SUPPLY SHOWED A BUILD UP OF DEBRIS (PRIMARILY DUST). THE POWER SUPPLY COMPONENTS SHOWED SIGNS OF THERMAL AND SMOKE DAMAGE. DAMAGE WITHIN THE POWER SUPPLY WAS LIMITED TO THE INDUCTOR AND SURROUNDING AREA. (B)(4), PROVIDED EXPONENT FAILURE ANALYSIS ASSOCIATES WITH ADD'L UNITS FOR FAILURE SIMULATION AND THE ANALYSIS IS ONGOING. PRELIMINARY WORK INDICATES THE CAUSE OF THE SMOKE WAS OVERHEATING OF THE INDUCTOR DUE TO A BUILD UP OF HEAT CAUSED BY DEBRIS IN THE POWER SUPPLY. INCIDENT SIMULATION CONDUCTED BY OTSUKA ELECTRONICS: OTSUKA ELECTRONICS IN (B)(4) CONDUCTED SIMULATION TESTING OF THE FAILURE BASED ON DESCRIPTION OF THE INCIDENT. TESTS WERE PERFORMED TO SIMULATE FAILURE IN NORMAL AND FAULT CONDITIONS. CONCLUSION: THERE IS NO RISK OF FIRE EVEN IF THE DEBRIS ACCUMULATED IN THE POWER SUPPLY SMOKES. ACCUMULATION OF DEBRIS IS IN PROPORTION TO THE USAGE HOURS AS A WHOLE AND DEGRADATION OF COMPONENTS IN THE POWER SUPPLY COULD POSSIBLY SMOKE. ADD'L INFO FROM IMPORTER REPORT: DATE OF THIS REPORT: (B)(4) 2007, (B)(6). INITIAL FEEDBACK AND CONFIRMATION: ACCORDING TO THE PRACTICE MGR, (B)(6) AND FIREMAN CALLED TO THE SCENE AT 8:41 AM, THERE ARE NO REPORTED INJURIES OR PROPERTY DAMAGE, AND NO FIRE (FLAMES) WERE OBSERVED DURING THE ENTIRE INCIDENT. VISUAL IMPRESSIONS AND EXAMINATION: NO VISUAL EVIDENCE OF FIRE OR SMOKE DAMAGE WAS OBSERVED, EITHER IN THE IMMEDIATE AREA WHERE THE UBIT-IR300 SPECTROPHOTOMETER WAS LOCATED, NOR IN THE ADJOINING HALLWAYS OR ROOMS OF THE CLINIC. NO VISUAL EVIDENCE OF FIRE OR SMOKE DAMAGE (BURNING, MELTING, CHARRING, SMOKE ACCUMULATION, ETC) WAS OBSERVED ON THE OUTSIDE OF THE UBIT-IR300 SPECTROPHOTOMETER UNIT, NOR ON THE INSIDE OF THE UNIT, AFTER (B)(6) REMOVED (ONLY) THE TOP COVER. OLFACTORY (SMELL) IMPRESSION: A DISTINCT SMELL, SIMILAR TO THAT OF BURNING ELECTRICAL INSULATION, WAS DETECTED BY BOTH (B)(6) AROUND, AND ESPECIALLY FROM WITHIN THE UBIT-IR300 SPECTROPHOTOMETER AFTER THE COVER WAS REMOVED. THE SMELL SEEMED TO BE STRONGEST WHEN SNIFFED WITHIN THE UNCOVERED UNIT. CAUSE OF THE COMPLAINT: IT WAS NOT POSSIBLE TO DETERMINE THE EXACT CAUSE OF THE COMPLAINT DURING THIS SITE VISIT. UBIT-IR300 SPECTROPHOTOMETER: SERIAL NO (B)(4) WAS SUBSEQUENTLY SHIPPED TO EXPONENT FAILURE ANALYSIS ASSOCIATES ((B)(4)) FOR FURTHER ANALYSIS AND CAUSE DETERMINATION. INVESTIGATION IS CONTINUING.

Description of Event or Problem · 1

ON THE MORNING OF (B)(6) 2007, THE UBIT-IR300 WAS IN WARM-UP PHASE AND THE STAFF NOTICED THAT IT WAS EMITTING SMOKE. THE BUILDING ALARM WAS ACTIVATED AND STAFF CALLED THE FIRE DEPT. THERE WERE NO INJURIES OR PROPERTY DAMAGE EXCEPT TO THE UBIT-IR300. MORE DETAILS OF THE INCIDENT WERE REPORTED IN THE IMPORTER REPORT NUMBER (B)(4). "STAFF WALKED INTO THE OFFICE AROUND 9:30 TO THE SMELL OF SMOKE. NOTICED THAT THE UBIT MACHINE WAS ON FIRE. FIRE ALARM WENT OFF; ENTIRE BLDG HAD TO BE EVACUATED. NO INJURIES." THE REPORTER CLARIFIED HER STATEMENT THAT THERE WAS NO FIRE AND THE SUBSEQUENT INVESTIGATION CONCLUDED THERE WAS SMOKE BUT NO FIRE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UBIT-IR300 INFRARED SPECTROPHOTOMETER JJQ (SPECTROPHOTOMETER) JJQ OTSUKA ELECTRONICS CO. LTD. UBIT-IR300 NA

Patients

Seq Age Sex Outcome Treatment
1