FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 2894946 · Received January 3, 2013

Report

Report Number
1416980-2013-00141
Event Type
Malfunction
Date Received
January 3, 2013
Date of Event
December 11, 2012
Report Date
December 11, 2012
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). COMPLAINT IS CONFIRMED BECAUSE IT WAS REPORTED DURING TROUBLE SHOOTING OF A LEAK ISSUE THAT THE CUSTOMER SWITCHED THE LEAKY SOLUTION BAG ONLY, WHICH IS A USE ERROR/POOR ASEPTIC TECHNIQUE. THE ROOT CAUSE WAS UNDETERMINED. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE USE ERROR IN THIS COMPLAINT. THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A SAMPLE EVALUATION WILL NOT BE CONDUCTED. THE LOT NUMBER WAS NOT PROVIDED, THEREFORE NO BATCH REVIEW COULD BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER ON (B)(4) 2012 FROM A PERITONEAL DIALYSIS (PD) PATIENT THAT ON (B)(6) 2012, THERE WAS A LEAK FROM THE SOLUTION BAG TUBING; THE PATIENT SUBSTITUTED THE BAG WITH ANOTHER ONE OF THE SAME LOT AND RESUMED THERAPY. IT IS NOT KNOWN IN WHICH STEP OF THERAPY THIS HAPPENED; NO SYMPTOMS OR CONSEQUENCES WERE REPORTED FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2230 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 HOME CHOICE