EON MINI
Report
- Report Number
- 1627487-2012-12814
- Event Type
- Injury
- Date Received
- December 20, 2012
- Date of Event
- November 27, 2012
- Report Date
- November 27, 2012
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-12192011-003-R
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
1627487-07262012-001-C. THIS IPG SERIAL NUMBER WAS INCLUDED IN FIELD ADVISORIES AND A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 4. REFERENCE MFR REPORTS: #1627487-2012-12815, #1627487-2012-12816 AND #1627487-2012-12817. IT WAS REPORTED THE PATIENT EXPERIENCED POCKET HEATING WHILE CHARGING. THE SJM REP REVIEWED THE CHARGING GUIDELINES WITH THE PATIENT. REPORTEDLY THE PHYSICIAN EXPLANTED THE SCS SYSTEM DUE TO PATIENT REQUEST. ON (B)(6) /2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3208616 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Required Intervention | SCS ANCHOR, MODEL 1192| IMPLANT DATE: |