FDA Adverse Event
Malfunction
Summary report: N
PROGRAMMING SOFTWARE
MDR report key: 2894845
·
Received January 3, 2013
Report
- Report Number
- 1644487-2013-00032
- Event Type
- Malfunction
- Date Received
- January 3, 2013
- Date of Event
- August 28, 2007
- Report Date
- December 4, 2012
- Manufacturer
- CYBERONICS INC
- Product Code
- MUZ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ANALYSIS OF PROGRAMMING HISTORY.
Description of Event or Problem · 1
DURING MANUFACTURER REVIEW OF THE PATIENT'S VNS PROGRAMMING HISTORY, IT WAS NOTED A FAULTED SYSTEMS TEST OCCURRED ON (B)(6) 2007 WHICH CHANGED THE VNS PROGRAMMED SETTINGS. THE SETTINGS WERE ALL CORRECTED EXCEPT FOR THE MAGNET ON TIME, WHICH WAS PREVIOUSLY AT 0MA AND WAS NOW AT 1MA. THE MAGNET OUTPUT CURRENT WAS LATER CORRECTED ON (B)(6) 2007.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2305 | PROGRAMMING SOFTWARE | PROGRAMMING SOFTWARE | MUZ | CYBERONICS INC | MODEL 250 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |