- Report Number
- 1825014-2008-00011
- Event Type
- Malfunction
- Date Received
- December 10, 2008
- Date of Event
- October 27, 2008
- Report Date
- December 8, 2008
- Manufacturer
- MIZUHO MEDICAL CO LTD
- Product Code
- GDC
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
WE RECENTLY RECEIVED REPORTS THAT SOME EXTREME POSITIONING / ARTICULATION MANEUVERS INITIATED DURING DAVINCI ROBOTIC PROCEDURES CAN POSSIBLY RESULT IN THE BACK SECTION COLLIDING WITH THE BASE OF THE 3600B TABLE. THIS EXTREME ARTICULATION CAN RESULT IN THE BACK SECTION BECOMING UNLOCKED. SECTIONS OF THE TABLE CAN CONFLICT WITH EACH OTHER; HOWEVER, IN THIS CASE SEVERAL STEPS ARE REQUIRED TO REPLICATE THE DISENGAGEMENT OF THE BACK SECTION. TO OUTLINE THIS EVENT, THE PT IS BEING PLACED IN A LITHOTOMY POSITION AND THE 3600B ARTICULATED INTO MAXIMUM TRENDELENBURG. WITH THE HEAD SECTION REMOVED, SOME SLIGHT BACK DOWN ADDED AND THE TABLE LONGITUDINAL MOTION ACTUATED TOWARD THE FOOT, THE RESULTING RATHER COMPLEX SERIES OF ARTICULATIONS AND MOVEMENTS IS AN UNLOCKING OF THE BACK SECTION. THE PROCEDURES REQUIRING THIS POSITION IS ARE DAVINCI PROSTATECTOMY AND HYSTERECTOMY CASES. THIS EVENT WAS NOT EASY TO REPRODUCE BUT WAS ACCOMPLISHED AND IS CURRENTLY BEING STUDIED. WE ARE RECOMMENDING THAT THOSE ACCOUNTS USING THE 3600B WITH THESE POSITIONAL REQUIREMENTS REVIEW THE 3600B OPERATOR'S MANUAL AND PRECAUTIONS OUTLINE IN THE SPECIAL USER ATTENTION SECTIONS.
DURING A DAVINCI ROBOTIC PROCEDURE AT (B)(6), THE 3600B SURGICAL TABLE WAS POSITIONED AS FOLLOWS: TRENDELENBURG ARTICULATION WAS DEPLOYED AT THE FULL 30 DEGREES OF HEAD DOWN, THE TABLE TOP SLIDE INITIATED TOWARD THE PT'S HEAD AND THE BACK DOWN FUNCTION ENGAGED SLIGHTLY, PRODUCING A DIRECT CONTACT POINT WITH THE TABLE BASE TOP COVER CORNERS AND THE BACK SECTION RELEASE LEVERS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SKYTRON | SURGICAL TABLE | GDC | MIZUHO MEDICAL CO LTD | 3600B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |