TANDEMHEART PUMP
Report
- Report Number
- 2531527-2008-00005
- Event Type
- Malfunction
- Date Received
- May 15, 2008
- Date of Event
- February 12, 2008
- Report Date
- May 12, 2008
- Manufacturer
- CARDIAC ASSIST, INC.
- Product Code
- KFM
- PMA / PMN Number
- K991783
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- NURSE
Narratives
EVAL SUMMARY: THE PUMP INVOLVED IN THE INCIDENT WAS RETURNED FOR EVAL. ALSO INCLUDED WAS THE INFUSION ASSEMBLY THAT WAS REPORTEDLY IN USE AT THE TIME OF THE INCIDENT. AN EVAL OF THE PUMP WAS CONDUCTED UPON RECEIPT WHICH CONFIRMED THAT THE PUMP WAS FUNCTIONAL AND MET THE PERFORMANCE REQUIREMENTS. NO DAMAGE OR DEFECTS WERE NOTED ON THE PUMP, EXCEPT FOR THE MODIFICATION TO THE PUMP INFUSION LINE THAT WAS PERFORMED BY THE CUSTOMER. INTERNAL ANALYSIS OF THE PUMP FOUND NO ABNORMAL WEAR OR DAMAGE. NO AIR LEAKS OR DAMAGE WERE NOTED ON THE PUMP INFUSION LINE OR LUER FITTING. A PRESSURE TEST OF THE INFUSION ASSEMBLY FOUND A LEAK AT THE DRIP CHAMBER, BUT NO DAMAGE OR LEAKS IN THE VICINITY OF TE BACTERIOLOGIC FILTER AS WAS REPORTED BY THE CUSTOMER. THE CAUSE OF AIR ENTRY INTO THE PUMP INFUSION LINE COULD NOT BE ESTABLISHED FROM THE ANALYSIS OF THE PUMP OR INFUSION ASSEMBLY.
AFTER APPROX 5 DAYS OF USE, AN INFUSION AIR BUBBLE ALARM OCCURRED ON THE CONTROLLER. SMALL AIR BUBBLES WERE OBSERVED ENTERING THE INFUSION ASSEMBLY AT THE BACTERIOLOGIC FILTER AND MIGRATING TOWARD THE PUMP. THE CLINICIANS ALSO REPORTED THAT A LOW FLOW ALARM OCCURRED ON THE CONTROLLER AND THE PUMP WAS OBSERVED TO BE CHATTERING. THE LOW FLOW ALARM WAS RESOLVED BY REORIENTING THE PT, REDUCING THE PUMP SPEED, AND ADDING VOLUME TO CORRECT PT FILLING ISSUES. THE CLINICIANS CLAMPED THE INFUSION LINE TO PREVENT FURTHER AIR MIGRATION IN THE INFUSION LINE AND CHANGED THE INFUSION ASSEMBLY. THE CLINICIANS CUT THE INFUSION LINE AT THE PUMP AND INSTALLED A STOPCOCK TO FACILITATE AIR REMOVAL FROM THE INFUSION LINE. THE PT WAS NOT IMPACTED BY THE INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TANDEMHEART PUMP | NON-ROLLER TYPE BYPASS PUMP | KFM | CARDIAC ASSIST, INC. | TANDEMHEART |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |