FDA Adverse Event Malfunction Summary report: N

TANDEMHEART PUMP

MDR report key: 2894505 · Received May 15, 2008

Report

Report Number
2531527-2008-00005
Event Type
Malfunction
Date Received
May 15, 2008
Date of Event
February 12, 2008
Report Date
May 12, 2008
Manufacturer
CARDIAC ASSIST, INC.
Product Code
KFM
PMA / PMN Number
K991783
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: THE PUMP INVOLVED IN THE INCIDENT WAS RETURNED FOR EVAL. ALSO INCLUDED WAS THE INFUSION ASSEMBLY THAT WAS REPORTEDLY IN USE AT THE TIME OF THE INCIDENT. AN EVAL OF THE PUMP WAS CONDUCTED UPON RECEIPT WHICH CONFIRMED THAT THE PUMP WAS FUNCTIONAL AND MET THE PERFORMANCE REQUIREMENTS. NO DAMAGE OR DEFECTS WERE NOTED ON THE PUMP, EXCEPT FOR THE MODIFICATION TO THE PUMP INFUSION LINE THAT WAS PERFORMED BY THE CUSTOMER. INTERNAL ANALYSIS OF THE PUMP FOUND NO ABNORMAL WEAR OR DAMAGE. NO AIR LEAKS OR DAMAGE WERE NOTED ON THE PUMP INFUSION LINE OR LUER FITTING. A PRESSURE TEST OF THE INFUSION ASSEMBLY FOUND A LEAK AT THE DRIP CHAMBER, BUT NO DAMAGE OR LEAKS IN THE VICINITY OF TE BACTERIOLOGIC FILTER AS WAS REPORTED BY THE CUSTOMER. THE CAUSE OF AIR ENTRY INTO THE PUMP INFUSION LINE COULD NOT BE ESTABLISHED FROM THE ANALYSIS OF THE PUMP OR INFUSION ASSEMBLY.

Description of Event or Problem · 1

AFTER APPROX 5 DAYS OF USE, AN INFUSION AIR BUBBLE ALARM OCCURRED ON THE CONTROLLER. SMALL AIR BUBBLES WERE OBSERVED ENTERING THE INFUSION ASSEMBLY AT THE BACTERIOLOGIC FILTER AND MIGRATING TOWARD THE PUMP. THE CLINICIANS ALSO REPORTED THAT A LOW FLOW ALARM OCCURRED ON THE CONTROLLER AND THE PUMP WAS OBSERVED TO BE CHATTERING. THE LOW FLOW ALARM WAS RESOLVED BY REORIENTING THE PT, REDUCING THE PUMP SPEED, AND ADDING VOLUME TO CORRECT PT FILLING ISSUES. THE CLINICIANS CLAMPED THE INFUSION LINE TO PREVENT FURTHER AIR MIGRATION IN THE INFUSION LINE AND CHANGED THE INFUSION ASSEMBLY. THE CLINICIANS CUT THE INFUSION LINE AT THE PUMP AND INSTALLED A STOPCOCK TO FACILITATE AIR REMOVAL FROM THE INFUSION LINE. THE PT WAS NOT IMPACTED BY THE INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TANDEMHEART PUMP NON-ROLLER TYPE BYPASS PUMP KFM CARDIAC ASSIST, INC. TANDEMHEART

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention