FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2894480 · Received December 27, 2012

Report

Report Number
8020893-2012-01546
Event Type
Malfunction
Date Received
December 27, 2012
Date of Event
November 28, 2012
Report Date
November 28, 2012
Manufacturer
COVIDIEN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD A BLANK SCREEN. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHIC USER INTERFACE CPU PCB. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK COVIDIEN 840

Patients

Seq Age Sex Outcome Treatment
1