FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BARIATRIC
MDR report key: 2894281
·
Received November 16, 2012
Report
- Report Number
- 1824206-2012-07502
- Event Type
- Malfunction
- Date Received
- November 16, 2012
- Date of Event
- October 26, 2012
- Report Date
- August 9, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECH FOUND THAT THE LATCH COVER WAS BENT PREVENTING THE RAIL FROM LOCKING IN THE UP POSITION. HE REPLACED THE LATCH COVER TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
INFO RECEIVED INDICATED THAT THE LEFT SIDERAIL WOULD NOT LATCH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE BARIATRIC | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |