XIA 3 TITANIUM TORQUE WRENCH
Report
- Report Number
- 9617544-2012-00607
- Event Type
- Injury
- Date Received
- December 27, 2012
- Date of Event
- November 28, 2012
- Report Date
- November 29, 2012
- Manufacturer
- STRYKER SPINE BORDEAUX
- Product Code
- NKB
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
METHOD: VISUAL INSPECTION, MANUFACTURING RECORD REVIEW, COMPLAINT HISTORY ANALYSIS, EXTERNAL ANALYSIS AND RISK ASSESSMENT. RESULTS: AFTER VISUAL INSPECTION THE TIP OF THE RETURNED DEVICE WAS FOUND TO BE BROKEN. MACHINING LINES CAN BE OBSERVED ON THE INSTRUMENT. MANUFACTURING RECORD REVIEW: NO ANOMALIES THAT COULD BE ASSOCIATED WITH THE BREAKAGE WERE REPORTED DURING THE MFG. COMPLAINT HISTORY ANALYSIS: 9 PREVIOUS RECORDS HAVE BEEN RECEIVED FOR HEX TIP BREAKAGE OF XIA 3 TORQUE WRENCH ON THE CURRENT DESIGN. EXTERNAL ANALYSIS PERFORMED ON TWO SIMILAR BROKEN DEVICES: ANALYSES SHOWS THAT BOTH DEVICES FAILED AS A RESULT OF SEMI-FRAGILE SUDDEN RUPTURE PROCESS INITIATED BY AN IMPORTANT NOTCH EFFECT. THE FAILURE INITIATED PRECISELY IN THE AREA BETWEEN THE HEXAGONAL EXTREMITY AND THE 8.5MM DIAMETER ZONE. NO ANOMALIES WERE FOUND ON HARDNESS MEASURES, MICROGRAPHIC STRUCTURE AND CHEMICAL COMPOSITION. RISK ASSESSMENT: REVISION SURGERY CONDUCTED TO REMOVE BROKEN FRAGMENT LEFT IN THE PATIENT 2 DAYS BEFORE WHEN THE INSTRUMENT BROKE. CONCLUSION: IT IS BELIEVED THAT THIS ISSUE IS MULTI-FACTORIAL. A NC (2012-043) HAS BEEN OPENED INTERNALLY IN ORDER TO ADDRESS THIS ISSUE.
ON (B)(6) 2012, XIA3 SURGERY WAS PERFORMED. ON (B)(6) 2012, THE OTHER PATIENT UNDERWENT THE SURGERY WITH THE XIA3. THE SURGEON TRIED TO USE THE TORQUE WRENCH, IT TURNED OUT THAT THE TIP OF THE TORQUE WRENCH HAS BROKEN. THEREFORE WHEN THE SURGEON CHECKED X-RAY AND MRI IMAGE OF THE PATIENT WHO PERFORMED AN OPERATION ON (B)(6) 2012, IT TURNED OUT THAT THE PIECE OF BREAKAGE OF THE TORQUE WRENCH REMAINS IN THE PATIENT'S BODY. ON (B)(6) 2012, THE OPERATION WHICH REMOVES THE PIECE OF BREAKAGE OF THE TORQUE WRENCH WAS PERFORMED. THE SURGEON REQUESTED THE INVESTIGATION (ELECTRON MICROSCOPY OF A FRACTURE CROSS SECTION) OF THE BROKEN TORQUE WRENCH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | XIA 3 TITANIUM TORQUE WRENCH | INSTRUMENT | NKB | STRYKER SPINE BORDEAUX | NA | 11E046 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR | Required Intervention |