FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2894103 · Received January 3, 2013

Report

Report Number
1644487-2013-00024
Event Type
Malfunction
Date Received
January 3, 2013
Date of Event
November 28, 2012
Report Date
December 6, 2012
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED ON (B)(6) 2013 WHEN IT WAS REPORTED THAT THE HANDHELD AND FLASHCARD WERE RECEIVED BY THE MANUFACTURER FOR PRODUCT ANALYSIS ON (B)(6)2013. PRODUCT ANALYSIS IS STILL UNDERWAY AND HAS NOT YET BEEN COMPLETED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED ON (B)(4) 2013 WHEN PRODUCT ANALYSIS WAS COMPLETED ON THE HANDHELD AND FLASHCARD. AN ANALYSIS WAS PERFORMED ON THE HANDHELD AND NO ANOMALIES ASSOCIATED WITH THE HANDHELD WERE NOTED DURING TESTING USING THE AC ADAPTER OR THE MAIN BATTERY WITH A FULL CHARGE. THE HANDHELD PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. AN ANALYSIS WAS PERFORMED ON THE RETURNED FLASHCARD AND THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO SPECIFICATIONS. DURING THE ANALYSIS, NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS.

Description of Event or Problem · 1

ON (B)(4) 2012 IT WAS REPORTED THAT THE PHYSICIAN'S HANDHELD WAS FREEZING ON VARIOUS SCREENS THE PREVIOUS WEDNESDAY. MULTIPLE HARD RESETS WERE PERFORMED BUT THE SCREEN WAS STILL FREEZING. ATTEMPTS FOR THE RETURN OF THE HANDHELD AND FLASHCARD WERE MADE BUT THE PRODUCTS HAVE NOT BEEN RETURNED FOR PRODUCT ANALYSIS TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
3648 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. MODEL 250 1073167

Patients

Seq Age Sex Outcome Treatment
1