FDA Adverse Event Injury Summary report: N

2.0 MM 2 X 8 T PLATE LOCKING

MDR report key: 2893945 · Received December 19, 2012

Report

Report Number
2027754-2012-00015
Event Type
Injury
Date Received
December 19, 2012
Date of Event
August 27, 2012
Report Date
December 18, 2012
Manufacturer
OSTEOMED
Product Code
HRS
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PLATE, PN 333-2006, WAS IMPLANTED ON (B)(6) 2012, DUE TO A 3RD DEGREE FRACTURE OF THE METACARPAL BONE. APPROX 10 WEEKS AFTER SURGERY, AN X-RAY REVEALED FRAGMENTS DUE TO THE BROKEN PLATE. ACCORDING TO THE DOCTOR, THE PT DID NOT HAVE ANY TYPE OF ACCIDENT OR PAIN BEFORE THE X-RAY. HE FOLLOWED ALL MEDICAL INSTRUCTIONS. ON (B)(6) 2012, THE DOCTOR REMOVED THE BROKEN PLATE AND PUT IN A NEW PLATE, PN 333-2006, AND A NEW SCREW. THEY STARTED THE "KINE" AND "FLEX" EXTENSION WITH DYNAMIC FERULA FLEXORA. AFTER THE SURGERY, THERE WAS SATISFACTORY EVOLUTION WITHOUT PAIN AVO DORSAL. THE PT IS IN GOOD CONDITION. THE RETURNED PLATE HAD A CLEAN MIDDLE BREAK IN THE 3RD HOLE. ACCORDING TO THE ENGINEER, A SCREW SHOULD HAVE PROBABLY BEEN PLACED PERPENDICULAR TO THE FRACTURE. IN ADDITION, PER OUR RESIDENT EXPERT (MEDICAL DOCTOR) BASED ON THE X-RAY, THE SCREW WAS PLACED TOO CLOSE TO THE FRACTURE LINE. BASED ON STANDARD AO FRACTURE FIXATION TECHNIQUES, A MINIMUM 1-2MM SPACING SHOULD EXIST BETWEEN THE FRACTURE SITE AND NEAREST SCREW. THERE WAS ONE OTHER COMPLAINT FOR THIS PART IN THE COMPLAINT DATABASE FROM AN INTERNATIONAL DOCTOR.

Description of Event or Problem · 1

THE PART WAS IMPLANTED ON (B)(6) 2012 AND EXPLANTED ON (B)(6) 2012, APPROX 2.5 MONTHS AFTER IMPLANTATION DUE. IT WAS NOTED ON THE POST - OP X-RAY THAT THE PLATE HAD BROKEN. THE PLATE WAS EXPLANTED APPROX 3 MONTHS AFTER IMPLANTATION AND ANOTHER PLATE WAS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 2.0 MM 2 X 8 T PLATE LOCKING BONE PLATE HRS OSTEOMED 333-2006 UNK

Patients

Seq Age Sex Outcome Treatment
1 31 YR Required Intervention