FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 2893913 · Received January 2, 2013

Report

Report Number
9611451-2013-00006
Event Type
Malfunction
Date Received
January 2, 2013
Date of Event
November 22, 2012
Report Date
December 7, 2012
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT MR290 CHAMBER IS CURRENTLY IN TRANSIT TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE COMPLAINT DEVICE AND COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290 CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) AND VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THE COMPLAINT CHAMBER DOME WAS CRACKED AT THE BAFFLE. A LOT CHECK REVEALED NO OTHER COMPLAINT OF THIS TYPE FOR LOT NUMBER 100620. CONCLUSION: WE WERE UNABLE TO DETERMINE THE ROOT CAUSE, ALTHOUGH IT IS KNOWN THAT PRESSURES PRODUCED IN EXCESS OF 80 CMH20 MAY AFFECT THE INTEGRITY OF THE CHAMBER. EVERY MR290 CHAMBER IS PRESSURE TESTED TO 8KPA (200CMH20) FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DOME DUE TO CRACKS AND OTHER CAUSES. ANY CHAMBER WHICH FAILS THIS TEST IS REJECTED. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. THE CUSTOMER HAS REPORTED THAT THE CRACK APPEARED AFTER 1 TO 2 DAYS OF USE, INDICATING THE FAULT OCCURRED POST PRODUCTION. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT AN MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS CRACKED AT THE DOME AFTER 1-2 DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT AN MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS CRACKED AT THE DOME AFTER 1-2 DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
554 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 100620

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening