ADULT INSPIRATORY HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00005
- Event Type
- Malfunction
- Date Received
- January 2, 2013
- Date of Event
- December 3, 2012
- Report Date
- December 6, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT RT205 BREATHING CIRCUIT IS CURRENTLY IN TRANSIT TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE COMPLAINT DEVICE AND COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE COMPLAINT RT205 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4). THE DEVICE WAS PRESSURE TESTED AND SUBMERGED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: THE PRESSURE TEST REVEALED THE PRESSURE DROP TO BE OUTSIDE OF SPECIFICATION FOR THIS PRODUCT. THE WATER BATH TEST IDENTIFIED THE LEAK TO BE AT THE CONNECTION BETWEEN THE LID AND BOWL OF THE WATER TRAP. CONCLUSION: THE LEAK WAS CAUSED BY A FAILURE IN THE SEAL BETWEEN THE WATER TRAP BOWL AND LID. ALL CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS ANY LEAK MUST HAVE DEVELOPED POST PRODUCTION DURING TRANSPORT, STORAGE OR USE. THE USER INSTRUCTIONS FOR RT205 ADULT BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARM."
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT AN RT205 ADULT BREATHING CIRCUIT DID NOT PASS THE SERVO VENTILATOR LEAK TEST. IT WAS FOUND THE THE DEVICE WAS LEAKING AIR AT THE CONNECTION BETWEEN THE WATER TRAP BOWL AND THE LID. THIS WAS FOUND PRIOR TO PATIENT USE.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT AN RT205 ADULT BREATHING CIRCUIT DID NOT PASS THE SERVO VENTILATOR LEAK TEST. IT WAS FOUND THE THE DEVICE WAS LEAKING AIR AT THE CONNECTION BETWEEN THE WATER TRAP BOWL AND THE LID. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 835 | ADULT INSPIRATORY HEATED BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LTD | RT205 | 120827 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |