FDA Adverse Event Malfunction Summary report: N

LARGE NEEDLE DRIVER

MDR report key: 2893620 · Received January 2, 2013

Report

Report Number
2955842-2013-00002
Event Type
Malfunction
Date Received
January 2, 2013
Date of Event
December 5, 2012
Report Date
December 5, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS FOUND WITH A FRAYED PITCH CABLE AT DISTAL IDLER LOCATION. THERE WAS NO DAMAGE FOUND ON PULLEY OR CLEVIS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CLEANING AND STERILIZATION, THE CUSTOMER NOTED THAT THE FRAYED CABLES ON THE LARGE NEEDLE DRIVER INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
436 LARGE NEEDLE DRIVER ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420006-06 M10120618 356

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI S SYSTEM INSTRUMENTS AND ACCESSORIES