FDA Adverse Event
Malfunction
Summary report: N
LARGE NEEDLE DRIVER
MDR report key: 2893620
·
Received January 2, 2013
Report
- Report Number
- 2955842-2013-00002
- Event Type
- Malfunction
- Date Received
- January 2, 2013
- Date of Event
- December 5, 2012
- Report Date
- December 5, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS FOUND WITH A FRAYED PITCH CABLE AT DISTAL IDLER LOCATION. THERE WAS NO DAMAGE FOUND ON PULLEY OR CLEVIS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING CLEANING AND STERILIZATION, THE CUSTOMER NOTED THAT THE FRAYED CABLES ON THE LARGE NEEDLE DRIVER INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 436 | LARGE NEEDLE DRIVER | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420006-06 | M10120618 356 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DAVINCI S SYSTEM INSTRUMENTS AND ACCESSORIES |