ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-00040
- Event Type
- Malfunction
- Date Received
- January 2, 2013
- Report Date
- December 7, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THERE IS A CRACK ALONG THE BATTERY COMPARTMENT. THE DOWN AND CONTRAST BUTTONS HAVE AN INTERMITTENT RESPONSE. NO VISIBLE DAMAGE TO KEYPAD. REMOVED KEYPAD AND FOUND CONTAMINATION UNDER ALL CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED CONTAMINATION UNDER BUTTON CONTACTS AND A BATTERY COMPARTMENT CRACK. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON (B)(4) 2012.