FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2892900 · Received January 2, 2013

Report

Report Number
2531779-2013-00040
Event Type
Malfunction
Date Received
January 2, 2013
Report Date
December 7, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THERE IS A CRACK ALONG THE BATTERY COMPARTMENT. THE DOWN AND CONTRAST BUTTONS HAVE AN INTERMITTENT RESPONSE. NO VISIBLE DAMAGE TO KEYPAD. REMOVED KEYPAD AND FOUND CONTAMINATION UNDER ALL CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED CONTAMINATION UNDER BUTTON CONTACTS AND A BATTERY COMPARTMENT CRACK. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON (B)(4) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1209 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 27 YR