CHARGING SYSTEM
Report
- Report Number
- 1627487-2012-02925
- Event Type
- Injury
- Date Received
- December 28, 2012
- Date of Event
- December 5, 2012
- Report Date
- December 6, 2012
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODUALTION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MANUFACTURER'S EVALUATION: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVALUATION: RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2012-02924. IT WAS REPORTED, THE PT EXPERIENCED AN EXTREME HEATING SENSATION AFTER 3-5 MINUTES OF CHARGING HIS IPG. THE PT STATED IT FELT LIKE THE HEAT WAS COMING FROM THE CHARGER ANTENNA AND NOT THE IPG SITE. HE ALSO REPORTED, HE EXPERIENCED INTERMITTENT CHARGING AS HIS CHARGER FREQUENTLY LOSES COMMUNICATION WITH HIS IPG. A REPLACEMENT CHARGING SYSTEM WAS SENT TO THE PT. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS AS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CHARGING SYSTEM | SCS CHARGING SYSTEM | LGW | ST. JUDE MEDICAL - NEUROMODUALTION | 3721 | 3865798 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Required Intervention | IMPLANT DATE:| SCS ANCHOR: MODEL 1192 (2)| IMPLANT DATE:| SCS LEAD: MODEL 3186 (2) |