CHARGING SYSTEM
Report
- Report Number
- 1627487-2012-15115
- Event Type
- Injury
- Date Received
- December 28, 2012
- Date of Event
- December 4, 2012
- Report Date
- December 4, 2012
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVAL CODES: RESULT: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMP TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REF MFR REPORT: 1627487-2012-15114. IT WAS REPORTED THE PT HAD NOT CHARGED HER SCS IPG IN MORE THAN 3 MONTHS. THE PT INDICATED THAT WHEN SHE DID CHARGE THE IPG SHE EXPERIENCED HEATING AT THE IPG POCKET SITE. THE SJM REP MET WITH THE PT AND CONFIRMED THE IPG WILL NO LONGER COMMUNICATE WITH EXTERNAL DEVICES. F/U INDICATED THAT SURGICAL INTERVENTION MAY BE TAKEN AT A LATER DATE TO ADDRESS THIS ISSUE. ON 08/01/2012, ST JUDE MEDICAL - NEUROMODULATION DIV., SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS ARE EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CHARGING SYSTEM | SCS CHARGING SYSTEM | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3721 | 2896353 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Required Intervention | SCS ANCHOR: MODEL 1194 (2)| IMPLANT DATE:| IMPLANT DATE:| SCS LEAD: MODEL 3186 (2) |