OCTROD
Report
- Report Number
- 1627487-2012-14295
- Event Type
- Injury
- Date Received
- December 28, 2012
- Date of Event
- December 5, 2012
- Report Date
- December 5, 2012
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVALUATION: METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: REVIEW OF THE DHR FOUND A NONCONFORMANCE RELATED TO THE PRODUCT LOT. HOWEVER, THE INDIVIDUAL AFFECTED DEVICE WAS REWORKED AND ALL DEVICES WITHIN THE LOT MET ACCEPTANCE CRITERIA. THEREFORE, THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE COMPLAINT. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
THE PATIENT HAS TWO LEADS IMPLANTED WITH THE SAME LOT NUMBER. IT WAS REPORTED THE PATIENT'S OCCIPITAL LEAD (OFF LABEL USE) IS ERODING THROUGH THE SKIN NEAR THE HIS TEMPLE. SURGICAL INTERVENTION IS PENDING TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OCTROD | SCS LEAD | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3186 | 3286154 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Required Intervention | SCS IPG: MODEL 3788| IMPLANT DATE: |