HEARTSINE SAMARITAN 300P AND PAD PAK
Report
- Report Number
- 3004123209-2012-00814
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Date of Event
- November 28, 2010
- Report Date
- December 14, 2012
- Manufacturer
- HEARTSINE TECHNOLOGIES, LTD
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PAD DEVICE WAS INSTALLED ON (B)(4) 2010 AND OPERATED SUCCESSFULLY UNTIL A FAILED SELF TEST ON (B)(4) 2010. THERE ARE SEVERAL MORE FAILED SELF TESTS DUE TO A LOW BATTERY. THE USER WAS ALERTED TO THE PROBLEM BY THE RED LED STATUS INDICATOR BEING ILLUMINATED AND AN AUDIBLE BEEP. ON INSPECTION OF THE PAD DEVICE, AN INCONSISTENT VOLTAGE MEASUREMENT WAS FOUND WHICH WOULD INDICATE A FAULT WITH THE POGO-PINS. . THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P ARE FOR MULTI-USE.
THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE THE STATUS INDICATOR WAS FLASHING RED AND THE DEVICE WAS EMITTING A LOW BATTERY WARNING. A DEVICE IN THIS FAULT MODE, IF LEFT UNDETECTED, COULD RESULT IN THE FAILURE OF THE DEVICE TO PERFORM AS INTENDED IF REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES, LTD | PAD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |