FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 2892639 · Received December 20, 2012

Report

Report Number
3004123209-2012-00814
Event Type
Malfunction
Date Received
December 20, 2012
Date of Event
November 28, 2010
Report Date
December 14, 2012
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K052465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PAD DEVICE WAS INSTALLED ON (B)(4) 2010 AND OPERATED SUCCESSFULLY UNTIL A FAILED SELF TEST ON (B)(4) 2010. THERE ARE SEVERAL MORE FAILED SELF TESTS DUE TO A LOW BATTERY. THE USER WAS ALERTED TO THE PROBLEM BY THE RED LED STATUS INDICATOR BEING ILLUMINATED AND AN AUDIBLE BEEP. ON INSPECTION OF THE PAD DEVICE, AN INCONSISTENT VOLTAGE MEASUREMENT WAS FOUND WHICH WOULD INDICATE A FAULT WITH THE POGO-PINS. . THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE THE STATUS INDICATOR WAS FLASHING RED AND THE DEVICE WAS EMITTING A LOW BATTERY WARNING. A DEVICE IN THIS FAULT MODE, IF LEFT UNDETECTED, COULD RESULT IN THE FAILURE OF THE DEVICE TO PERFORM AS INTENDED IF REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES, LTD PAD

Patients

Seq Age Sex Outcome Treatment
1