FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 2892549 · Received December 20, 2012

Report

Report Number
3004123209-2012-00564
Event Type
Malfunction
Date Received
December 20, 2012
Date of Event
September 30, 2011
Report Date
November 26, 2012
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K052465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PAD DEVICE WAS INSTALLED ON (B)(6) 2010 AND OPERATED SUCCESSFULLY UNTIL (B)(6) 2011. AFTER THIS DATE THERE ARE MULTIPLE EVENTS ON THE SAME DAY WHICH WOULD INDICATED THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY. THE DEVICE WAS ALSO EXPOSED TO EXTREME TEMPERATURES OF 50 DEGREES C AND MORE. PREVIOUS EXPERIENCE HAS TOLD US THAT THE REPORTED SYMPTOMS CAN BE CAUSED BY A PROBLEM WITH THE DEVICE MEMBRANE. STORING THE PAD DEVICE AND PAD PAKS IN INADEQUATE CONDITIONS WHERE IT CAN BE EXPOSED TO ADVERSE CONDITIONS CAN ALSO AFFECT THE MEMBRANE. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED, COULD RESULT IN THE BATTERY BECOMING DEPLETED.