FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 2892529
·
Received November 26, 2012
Report
- Report Number
- 1824206-2012-07689
- Event Type
- Malfunction
- Date Received
- November 26, 2012
- Date of Event
- November 6, 2012
- Report Date
- November 6, 2012
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THE PROBLEM WAS WITH THE RIGHT USER CONTROL BOARD AND CABLE. HE REPLACED THE RIGHT UCM BOARD AND CABLE TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATED THE HEAD SECTION AND HI-LO RAISED WITH NO SWITCHES BEING DEPRESSED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE BED | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |