FDA Adverse Event
Malfunction
Summary report: N
HEARTSINE SAMARITAN 300P AND PAD PAK
MDR report key: 2892407
·
Received November 30, 2012
Report
- Report Number
- 3004123209-2012-00583
- Event Type
- Malfunction
- Date Received
- November 30, 2012
- Date of Event
- November 11, 2012
- Report Date
- November 26, 2012
- Manufacturer
- HEARTSINE TECHNOLOGIES, LTD
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE PAD DEVICE WAS INSTALLED ON (B)(6) 2009 AND OPERATED SUCCESSFULLY UNTIL (B)(6) /2011. AFTER THIS DATE, THERE ARE MULTIPLE EVENTS ON THE SAME DAY WHICH WOULD INDICATE THE DEVICE WAS SWITCHING ITSELF ON AUTOMATICALLY. THE PROBLEM WITH THE DEVICE WAS ATTRIBUTED TO A FAULT IN THE MEMBRANE. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE, BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE, WHICH IS WHY THE "UNKNOWN" BOX HAS BEEN CHECKED IN THIS REPORT.
Description of Event or Problem · 1
THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE IT WAS SWITCHING ITSELF ON AUTOMATICALLY. A DEVICE SWITCHING ITSELF ON AUTOMATICALLY, IF LEFT UNDETECTED, COULD RESULT IN THE BATTERY BECOMING DEPLETED.