DEXTRUS 4136
Report
- Report Number
- 1028232-2012-03296
- Event Type
- Injury
- Date Received
- December 18, 2012
- Date of Event
- October 2, 2012
- Report Date
- December 4, 2012
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVZ
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MFG PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MFG PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MFG.
BOSTON SCIENTIFIC RECEIVED INFO THAT THIS RIGHT ATRIAL LEAD EXHIBITED HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS ALONG WITH NOISE. AN INVASIVE PROCEDURE WAS PERFORMED TO REVISE THE LEAD. WHEN THE POCKET WAS OPENED IT WAS NOTED THE RA LEAD WAS NOT FULLY INSERTED INTO THE DEVICE HEADER. THE LEAD WAS REINSERTED INTO THE DEVICE HEADER AND THE SET SCREWS WERE SECURED WHICH RESOLVED THE CLINICAL OBSERVATION. NO ADVERSE PT EFFECTS WERE REPORTED. RECORDS INDICATE THIS DEVICE AND LEAD REMAIN IN SERVICE. AS NO FURTHER INFO CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFO BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DEXTRUS 4136 | PACER LEAD | NVZ | BIOTRONIK SE & CO. KG | 358754 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |