ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-00003
- Event Type
- Malfunction
- Date Received
- January 2, 2013
- Report Date
- December 21, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION (B)(4) 2013- DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE PEELING AT THE OK BUTTON. A DAMAGED KEYPAD WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACTS NEGATIVELY IMPACTING THE BUTTON FUNCTION. THE DAMAGED KEYPAD IS OBVIOUS AND DETECTABLE AND SHOULD ALERT THE USER TO DISCONTINUE USE OF THE PUMP. DURING TESTING, UP, DOWN, AND CONTRAST BUTTONS WERE INTERMITTENTLY UNRESPONSIVE; THE OK BUTTON RESPONDED APPROPRIATELY. DURING INVESTIGATION, EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A CRACKED BATTERY COMPARTMENT AND A DISCOLORED/FADED DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE USER GUIDE WARNS THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND/OR THE WATERPROOF FEATURE OF THE PUMP.
THE PATIENT CONTACTED ANIMAS ON (B)(4) 2012 REPORTING THAT THE OK BUTTON WAS INTERMITTENTLY UNRESPONSIVE WITH THE KEYPAD PEELING. THE PATIENT STATED THE INTERMITTENTLY UNRESPONSIVE BUTTON STARTED A FEW MONTHS AGO AND THE KEYPAD STARTED TO PEEL A WEEK AGO. THE PATIENT DENIED TRAUMA AND MOISTURE TO THE PUMP. THE PATIENT REPORTEDLY WORE THE PUMP IN A POCKET OR ON A BELT CLIP AND CLEANS THE PUMP WITH A DAMP TOWEL. THERE IS NO REPORTED ADVERSE EVENT WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON THE ALLEGATION OF THE KEYPAD RESPONSE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1702 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR |